The YSI TruLab line includes the 1110 (single channel), 1310 and 1310P (single channel) and 1320 and 1320P (dual channel)
instruments providing easy-to-use and calibrate instruments ideal for the laboratory.
Measure pH, mV (ORP), ISE and temperature accurately in the laboratory with the YSI pH 1320 or 1320P with built-in
printer. The TruLab pH/ISE 1320 benchtop meter is the ideal laboratory instrument for reliable measurements with
automatic documentation for GLP compliance.
The 1320 has internal memory. Stored data can be sent to a PC via USB or to an optional built-in printer available on the 1320P.
Features:
Parameters:
Benchtop | Yes |
---|---|
Buffer Recognition | Yes |
Certifications | cETLus, electrical protective class III, CE, IP43 |
Display | Backlit graphic display |
Equipment used with | TruLab analog pH/mV/ISE/temp probes |
GLP Compliance | Yes |
Graphic Display | Backlit graphic display |
Logging Capabilities | Yes |
Measurement Method / Channels | 2 Channel |
Measurement Range |
Ranges Resolutions Accuracy (+1 digit) pH -2.0 to 20.0; -2.00 to 20.00; -2.000 to 19.999 0.1; 0.01; 0.001 ±0.1; ±0.01; ±0.005 (temp of test sample 15 to 35°C; within a range of ±2 pH of the calibration point) mV -1200 to 1200.0; -2500 to 2500 0.1; 1.0 ±0.3; ±1.0 (temp of test sample 15 to 35°C) ISE (mg/l, mg/kg, mmol/l, ppm, %) (µmol/l) 0.000 to 9.999; 10.00 to 99.99; 100.0 to 999.9; 1000 to 999999 0.000 to 9.999; 10.00 to 99.99; 100.0 to 999.9; 1000 to 9999 0.001; 0.01; 0.1; 1 Temperature -5 to 105°C (23.0 to 221°F) 0.1 ±0.1 |
Memory | 5,000 data sets auto logging, 500 data sets manual logging |
Operating pH Range | -2.00 to 20.00, -2.0 to 20.0, -2.000 to 19.999 |
Operating Temperature | -5 to 105°C, 23.0 to 221°F |
Parameters Measured |
pH, ORP (mV), Ammonia, Ammonium, Bromide, Cadmium, Calcium, Chloride, Copper, Cyanide, Fluoride, Iodide, Lead, Nitrate, Potassium, Silver/Sulfide, Sodium, Temperature |
Power |
approximately 150 hours with illumination or 1000 without, 4 alkaline AA batteries fully charged or included universal power supply |
Sampling | Yes |
Storage Temperature | -25 to 65°C, -13 to 149°F |
Unit of Measure |
Sample addition, Known addition (double), Known subtraction, ISE Measurements: Blank value correction, increment methods: Known addition (single), Sample subtraction |
User Calibratable | Yes |
User Replaceable | Yes |
Warranty | 3 years |